Particle Size Range

5 nm – 2.5 μm diameter

Sample Flow Rate

0.3 L/min

Concentration Range

<1/cc – 2.5 x 10^5/cc

Averaging Time

1 second to 30 minutes, user selectable

Precision

Concentration measurement precision within 10% for co-located units sampling the same aerosol

Working Fluid

Water (distilled or cleaner)

Wick Sensor

Provides direct reading of wick moisture level

Orientation

Can be operated for extended periods of time in any orientation

Data Logging

On-board non-volatile data storage. 262144 records; approximately 15 days of 5-second averaged data

Communications

Serial Type B USB and RS-232 9-pin DIN; digital pulse BNC

Output Data

Timestamp, concentration, raw counts, live time, clock time, temperature readings, input T/RH, flow, pressure, status flags

Dimensions

16 cm (D) x 20 cm (W) x 23 cm (H) (8.0 x 6.5 x 9.0 inches)

Weight

2.5 Kg (excludes power supply)

Power supply and line cord weight adds 0.4 Kg

Power Requirements

100-240 VAC with power supply; or 12.0 VDC; 16 Watts

Warranty

One year

Manufactured in the U.S.A. by Aerosol Dynamics Inc. U.S. Patents #6712881 and #8801838; German patent #10392241. Other patents pending.

Hering, S. V., Spielman, S. R., Lewis, G. L. (2014) Moderated, water-based condensational growth of particles in a laminar flow. Aerosol Science and Technology 48:401-40.